Description
COD . Ammonia Nitrogen Dual Parameter Rapid Tester 5B-3C(V8)
1 >Product Introduction
Fully based on national regulations "Determination of water content of chemical oxygen demand rapid digestion spectrophotometry (HJ / T399-2007)" and "Determination of ammonia nitrogen in Water Quality Nessler reagent spectrophotometric method (HJ535-2009)" design and manufacture, applicable to many domestic and foreign testing standards.
2 >Functional characteristics
(01)Quick and accurate determination of chemical oxygen demand (COD) in surface water, reclaimed water, municipal wastewater and industrial wastewater. Concentration direct reading.
(05)Self-calibration function of the instrument can be calculated and stored according to the standard sample, without manual production of curves.
(06)Large and small font display mode is free to switch, showing clearer data and more detailed parameters.
3>Technical Parameters
Item | COD high range | COD low range |
Range | 20-10000mg/L(subsection) | 2-150mg/L(subsection) |
Measurement accuracy | COD<50mg/L,error≤±10%; | Error≤±10% |
COD>50mg/L,error≤±5% | ||
limits of detection | 0.1mg/L | 0.1mg/L |
Determination time | 20min | 20min |
Batch processing | 16 | 16 |
Repeatability | ≤±5% | |
Lamp life | 100 thousands hours | |
Optical stability | ≤±0.005A/20min | |
Anti-chlorine interference |
<1000mg/L no influence ;
<100000mg/L Optional
|
|
Colorimetric method | Cuvette/Tube | |
Data storage | 12000 | |
Curve data | 180 | |
Display mode | LCD(Resolution 320*240) | |
Communication interface | USB /Infar-red (optional) | |
Power supply | AC220V±10%/50Hz |